Chapter 6
|
Troubleshooting
6-2
|
Precision
Checklist
Bath will not start or shuts down
Check display for error messages.
Ensure
wasn't accidently pressed.
Ensure the circuit protector is in the on ( I ) position.
Check the line cord connection to your power supply and at the bath.
Make sure supply voltage is connected and matches the bath's nameplate rating ±10%.
Restart the bath.
No display
Pressing
should return the display.
Cycle the bath's circuit protector.
Inadequate temperature control
Verify the set point.
Low-end temperature set points require supplemental cooling.
Ensure bath installation complies with the site requirements in Chapter 3.
Make sure supply voltage matches bath's nameplate rating ±10%.
Please contact Thermo Fisher Scientific Sales Service and Customer Support if you need any
additional information, see inside cover for contact instructions.