• Do not for any reason occlude the external
access hole of pressure relief valve.
• Do not occlude the gas escape port to avoid
that pressure achieved into the gas side
compartment becomes higher than pressure
into the blood site compartment.
VENOUS LINE: Connect a 3/8" venous line to the
port on the venous reservoir marked "VENOUS
RETURN".
The Venous Return port can be rotated through
150° to find the most convenient position for the
venous tube.
ASPIRATION LINES: After removing the
protective caps from the filtered inlet ports
on the top of the cardiotomy/venous reservoir
(composed of 6 x 1/42 inlets, 1 x 3/8" inlet, 2 x
female luer lock and 2 x POS lock reducible to
female luer lock) connect the aspiration lines end
and rotate the turret orienting the filtered inlets
towards the suction pumps.
ARTERIAL LINE: Remove the red cap on
the arterial outlet of the oxygenator marked
"ARTERIAL OUTLET" and connect a 3/8" line.
PUMP LINE: The pump segment must be fitted
between the venous reservoir outlet port and the
oxygenator venous inlet port taking account of
the pump rotation direction.
CAUTION
CARDIOPLEGIA LINE: If oxygenated blood is
necessary for blood cardioplegia, connect the
TRILLY PAEDIATRIC AF cardioplegia blood line
to the cardioplegia circuit.
6) SAMPLING MANIFOLD
The arterial/venous sampling manifold is
positioned on the venous reservoir lid and is
already connected to the sampling points.
If you wish to take samples from a more remote
position, remove the manifold from the fastening
guide, fit it on the dedicated holder and unwind
the tubing provided. In this position, arterial/
venous sampling can be performed in a radius of
about 1 meter.
WARNINGS:
• The female luer arterial sampling port,
located over the arterial outlet port, pre-
connected with sampling manifold, cannot be
used as outlet pressure measurement port
due to the presence of integrated one-way
valve.
• Blood pressure monitoring can be performed
by connecting to the luer lock located under
the arterial port.
• Male luer locks not supplied with EUROSETS
products may damage the one-way valve
positioned in the oxygenator arterial
sampling luer lock.
7) CONNECTING THE TEMPERATURE PROBES
The arterial temperature probe port is
positioned near the arterial outlet, while the
venous temperature probe port is opposite to the
oxygenator venous inlet.
8) CLOSING THE PURGE AND CARDIOPLEGIA
LINES
Connect the purge line to the turret; close the
clamp on the cardioplegia line and the clamp on
the purge line.
Connect the cardioplegia line to the cardioplegia
kit.
9) CONNECTING THE GAS LINE
Remove the green cap from the gas inlet port
marked "GAS INLET" and connect the 1/4" gas
line.
Ensure that the gas supply is from a suitable air/
oxygen mixer such as Sechrist or a system with
similar technical features.
The special configuration of the "GAS ESCAPE"
port prevents potential occlusion of the gas
outlet.
A capnograph can be attached to this port.
CAUTION
The "GAS ESCAPE" system has been designed
to avoid any possible risk of occluding the gas
outlet, which would cause immediate passage of
air into the blood compartment.
The user should check for occlusions in tubing
during set up.
PRIMING AND RECIRCULATION PROCEDURE
WARNINGS
• Do not apply the vacuum during this procedure.
• Do not use alcohol-based priming solution, as
this would compromise proper functioning of
the oxygenator module.
1) KEEP THE GAS FLOW OFF
2) KEEP THE OXYGENATOR PURGE LINE CLOSED
Check that the purge line clamp is closed.
3) CLOSE VENOUS AND ARTERIAL LINES