1-3 Electrical Specifications
Accuracy is ± (% reading + number of
digits) at 23°C ± 5°C < 80%RH.
Temperature coefficient :
Add 0.2 x (Specified accuracy) / °C, < 18°C,
> 28°C .
Voltage
Function
Range
0~400.0Vrms
±(1.0%+ 5 digits)
V
400~600Vrms
50Hz ~ 500Hz *1
0~400.0 V
V
±(0.7% + 2 digits)
400~600 V
*1: For readings of less than 15% of full
scale (4000 digits), add 4 digits to reading.
The settling time will be longer.
Overload protection: 600 Vrms
Input impedance: 1MΩ // less than 100pF.
AC Conversion Type:
AC Conversions are ac-coupled, true rms
responding, calibrated to the rms value of
a sine-wave input. Accuracies are given for
sine-wave at full-scale and non-sine wave
below half-scale. For non sine-wave add
the following Crest Factor corrections:
For Crest Factor of 1.4 to 2.0, add 1.0% to
accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to
accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to
accuracy.
CF
≦
2 600V//600A
11
Accuracy