14. Declaración de conformidad CEM
The device has been tested and homologated in accordance with EN 60601-1-2 for
EMC. This does not guarantee in any way that this device will not be affected by
electromagnetic interference. Avoid using the device in high electromagnetic envi-
ronment.
14.1 Technical Description Concerning Electromagnetic Emission
Table 1: Guidance & declaration - electromagnetic emissions
The model CanalPro X-Move is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the model CanalPro X-Move
should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR11
Harmonic
emissions
lEC 61000-3-2
Voltage fluctuations
/ flicker emissions
lEC 61000-3-3
14.2 Technical Description Concerning Electromagnetic Immunity
Table 2: Guidance & Declaration - electromagnetic immunity
The model CanalPro X-Move is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the model CanalPro X-Move
should assure that It is used in such an environment.
Immunity test
Electrostatic
discharge
(ESD)
lEC 61000-4-2
32
Compliance
Group 1
Class B
Class A
Complies
IEC 60601
test level
±8kV contact
±2, ±4, ±8,
±15kV air
Electromagnetic environment - guidance
The model CanalPro X-Move uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The model CanalPro X-Move is suitable for
used in all establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply net-
work that supplies buildings used for domes-
tic purposes.
Compliance level
±8kV contact
±2, ±4, ±8, ±15kV
air
Electromagnetic environ-
ment - guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humid-
ity should be at least 30 %.