Malfunction
Er:X:16
...
is displayed?
... when the Enter key
(send/print) is pressed
Er:X:71
and
played?
... when the Enter key
(send/print) is pressed
Er:X:72
and
played?

10.MAINTENANCE/RECALIBRATION

10.1 Information about maintenance and recalibration

64 •
Oscillation of the scales has occurred, preventing
determination of the zero point.
- Restart scales
Data transmission not possible, wireless module is
disabled.
is dis-
- Activate wireless module (see "Activate
Data transmission not possible, no wireless group
set up.
- Set up wireless group (see "Set up wireless
is dis-
We recommend having maintenance performed prior
to recalibration of the scales.
Ensure that recalibration is carried out by authorised
persons in accordance with the national legal require-
ments. The year in which initial calibration was per-
formed can be found behind the CE mark on the rating
plate above the number of the notified body 0109
(Hesse Weights & Measures Office).
Recalibration is required in all cases if one or more cali-
bration seals have been damaged or the content of the
calibration counter does not match the number on the
valid calibration counter sticker.
Cause/Remedy
wireless module (SYS)" on page 58).
group (Lrn)" on page 59).
ATTENTION!
Incorrect measurement due to improper
maintenance
– Make sure that maintenance and repair are
only carried out by an authorised service
partner.
– You can find your local service partner at