Chapter VII
Dissolved Oxygen Technique
Dissolved Oxygen Setup
The following table describes dissolved oxygen setup .
Description
Dissolved Oxygen
% Saturation Resolution
Dissolved Oxygen
Concentration Resolution
Dissolved Oxygen
Barometric Pressure
Compensation Selection
Dissolved Oxygen Manual
Barometric Pressure
Setting
Dissolved Oxygen Salinity
Correction Selection
Dissolved Oxygen Manual
Salinity Correction Factor
Dissolved Oxygen
Calibration Type Selection
Thermo Scientific Orion Star
Display
Range of alues
1, 0.1
dO
rES
DO % saturation
0.1
resolution
dO
0.1, 0.01
rES
DO concentration
0.01
"mg/L" resolution
dO
AUtO, mAn
bAr
Selects internal barometer
AUtO
or manual pressure
450.0 - 850.0
dO
PrES
Manual pressure
760.0
compensation value
dO
AUtO, mAn
SAL
Selects method of salinity
AUtO
correction; meters with
conductivity function only
0 - 45
dO
SALF
Manual salinity correction
0
factor
dO
AIr, H2O, mAn, SEt0
CALt
AIr = Water sat. air
AIr
H2O = Air sat. water
mAn = Manual
SEt0 = Zero point
Series Meter
Method
Default
Specific
alue
Yes
0.1
Yes
0.01
Yes
AUtO
Yes
760.0
Yes
AUtO
0
Yes
Yes
AIr