10. I L VA P O R I Z Z AT O R E M I G H T Y M E D I C
1) Field strengths from fixed transmit-
ters, such as base stations for radio
(cellular/cordless)
land mobile radios, amateur radio,
AM and FM radio broadcast and TV
broadcast cannot be predicted theo-
retically with accuracy. To assess the
electromagnetic environment due to
fixed RF transmitters, an electromag-
netic site survey should be considered.
If the measured field strength in the
location in which the MIGHT Y MED-
IC Vaporizer is used exceeds the
Recommended Separation Distances between portable and mobile RF Com-
munications Equipment and the MIGHTY MEDIC Vaporizer
The MIGHT Y MEDIC Vaporizer is intended for use in an electromagnetic en-
vironment in which radiated RF disturbances are controlled. The customer or
the user of the MIGHT Y MEDIC Vaporizer can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mo-
bile RF communications equipment (transmitters) and the MIGHT Y MEDIC
Vaporizer as recommended below, according to the maximum output power
of the communications equipment.
Rated maximum
output power of
transmitter
[W]
0,01
0,1
1
10
100
telephones
and
Separation distance according to frequency of
150 kHz – 80 MHz
d = 1,2 √P
0,12
0,38
1,2
3,8
12
applicable RF compliance level above,
the MIGHT Y MEDIC Vaporizer should
be observed to verify normal opera-
tion. If abnormal performance is ob-
served, additional measures may be
necessary, such as re-orienting or
relocating the MIGHT Y MEDIC Va-
porizer.
2) Over the frequency range 150 kHz
to 80 MHz, field strengths should be
less than 3 V/m.
transmitter [m]
80 MHz – 800 MHz
d = 1,2 √P
0,12
0,38
1,2
3,8
12
248
800 MHz – 2,5 GHz
d = 2,3 √P
0,23
0,73
2,3
7,3
23