Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MIGHTY MEDIC Vaporizer is intended for use in the electromagnetic envi-
ronment specified below. The customer or the user of the MIGHTY MEDIC
Vaporizer should assure that it is used in such an environment.
Immunity
test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601
test level
± 6 kV contact
discharge
± 8 kV air
discharge
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Compliance level Electromagnetic environ-
± 6 kV contact
discharge
± 8 kV air
discharge
± 2 kV for power
supply lines
not applicable:
input/output
lines are
non-existent
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
94
ment – guidance
Floors should be wood,
concrete or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Mains power quality
should be that of a
typical residential,
commercial or hospital
environment.
Mains power quality
should be that of a
typical residential,
commercial or hospital
environment.