Guidance and Manufacturer's Declaration (Electromagnetic Emissions)
Emissions Test
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Guidance and Manufacturer's Declaration (Electromagnetic Immunity)
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / burst
4
Compliance
Group 1
Class B
N/A
N/A
IEC 60601
Test level
± 2 kV, ± 4 kV,
± 8 kV, contact
± 8 kV, ± 15 kV, air
N/A
Electromagnetic Environment (Guidance)
Uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Is suitable for use in all establishments, including domestic
establishments and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
N/A
N/A
Compliance Level
± 2 kV, ± 4 kV,
± 8 kV, contact
± 8 kV, ± 15 kV, air
N/A
Electromagnetic Environment
(Guidance)
Floors should be wood, concrete,
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
N/A