Processing
5.1
Basic Information on Processing
The operator/user is responsible for cleaning and disinfecting the device. It is essential that
national regulations and restrictions be observed.
Danger of severe bodily injuries from damaged maXium
accessories!
A non-functional device or defective accessories can endanger the patient or user, as well as
affect the intended functions of the device.
•
To avoid incompatibilities and unsafe operation, only electrosurgical accessories and
medical-electrical devices by Gebrüder Martin may be selected and used.
The device and its accessories must always be kept in flawless working condition.
•
Intactness of the insulation, cleanliness and integrity must be visually ascertained.
•
The cart and its accessories must always be kept in flawless working condition.
The note regarding regular control of accessory equipment is intended for the user: Intact
•
insulation, cleanliness and potential damage must be visually ascertained (exclusion of
first errors).
•
If the cart does not work properly and/or cannot be operated safely, it must be marked as
non-operational and put out of operation.
Units and accessories unfit for further use are to be scrapped.
•
®
The maXium
smart Cart must not be sterilized.
•
•
When handling cleaning agents and disinfectants, even during spraying on no liquid may
enter the electrical connections of the maXium
Introducing shunt-creating substances into the maXium
•
shock.
Revision 1
All manuals and user guides at all-guides.com
®
smart Cart and damaged
®
smart Cart.
®
smart Cart can result in electric
43