The screen does not show anything :
Check that the device has enough battery
The screen displays body temperature less than to 86°F / 30°C:
To measure a body temperature, the BODY mode must be selected. If you are on SURFACE TEMP
mode, the 86°F / 30°C temperature displayed is showing the external temperature that your
body releases.
Anyone with continued BODY temperature readings less than 86°F / 30°C should seek
emergency medical help immediately.
A body temperature below 90 °F / 32°C is life threatening, and when it drops to 86 °F / 30°C or
lower, a person may slip into a coma.
As always, if you have any doubts, seek care from a physician.
The screen displays the message HI: when using the MyThermo BW-CX10, the message HI can
show on the screen. The analysis is above the measurement range selected, either higher than
109.4°F / 43°C in BODY Mode or higher than 199.9°F / 93.3°C in SURFACE TEMP Mode.
The screen displays the message LO: when using the MyThermo BW-CX10, the message Lo
can show on the screen. The temperature analyzed is under the measuring range selected,
either less than 86°F / 30°C in BODY Mode or less than 32°F / 0°C in SURFACE TEMP Mode.
This message displays in various cases – please find below a list of the main cases:
Temperature reading hampered by hair,
perspiration.
Temperature hampered by any air flow/
drafts.
Temperature readings too close together,
MyThermo did not have the chance to reboot
itself.
The measuring distance is too far.
Bluetooth connection
If you have trouble with the Bluetooth connection:
- Check that your are in range of the Bluetooth accessory which you are trying to pair with
e
MyThermo.
- Make sure that your Bluetooth accessory is turned on and fully charged or connected to power.
- Make sure that you are in Body Mode on your MyThermo. You cannot send Surface temp
measurement or Room measurement to your Bluetooth accessory.
Causes of Lo Message
BewellConnect - BW-CX10 - User Manual - 042017v3
Solutions
Make sure that there is no obstruction prior to
taking a temperature.
Make sure there is no air flow/drafts, as this
could interfere with the infrared system.
Wait a minimum of 15 seconds between two
readings – 1 minute pause is advised.
Please respect the measuring distance
(between 1 in and 2 in / 2 to 5 cm).