5.3. AC VOLTAGE MEASUREMENT
The maximum input AC Voltage is 600Vrms. Do not measure voltages
exceeding the limits given in this manual. Exceeding these limits could
result in electrical shocks to the user and damage to the instrument.
Fig. 5: Use of the instrument for AC Voltage measurement
1. Select the position V
2. Insert the red cable into input terminal VHz%
terminal COM
3. Position the test leads in the desired points of the circuit to be measured (see Fig. 5).
The display shows the value of voltage.
4. If the symbol "O.L" is displayed, this indicates overload status.
5. For use the HOLD, RANGE and MAX MIN functions, please refer to § 4.2.
CAUTION
Hz%
EN – 10
and the black cable into input
HT3013