NOTICE
► Loss of default settings
If the battery replacement process takes longer than 5 minutes,
the date and time must be reset. A measurement is only possible
with the time set.
4.3
Carrying out an automatic breath sample
A new mouthpiece must be inserted for each new person who
provides a breath test.
Do not allow any high alcohol concentrations on the sensor, e.g., by
rinsing the mouth with a high proof alcohol shortly before the
measurement, as this will shorten the service life of the sensor.
Procedure
1. Remove the mouthpiece (order no. 68 10 690 "Standard",
68 11 055 "with exhalation non-return valve") from the packaging.
For hygiene reasons, leave the protective wrapping around the
mouth contact area until the mouthpiece is firmly attached to the
device.
2. Attach the mouthpiece at an angle to the mouthpiece receptacle
and press it into the receptacle (see Fold-Out Page, figure B, C).
The fluted markings (press-on surface) on the mouthpiece face
upwards while this is done.
 The mouthpiece audibly snaps into place.
3. Remove all packaging.
Keep the
screen is displayed. An optional self-test is carried out. The
following is shown on the display for approx. 4 seconds:
● READY
 The device is ready to measure.
Dräger Alcotest 6000
button pressed for approx. 1 second until the startup
|
Use
en
11