CAUTION
Do not mount a maintenance Tip (V10-S or V-P10) to the Tips Holder as the holder lid cannot be fully closed. Use a
autoclaving pouch for autoclave sterilizing of maintenance Tips.
8-2-3 Autoclave
Autoclave up to max. 135˚C.
Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C.
Keep the sterilization cassette to keep it clean until you use it.
* Sterilization at 121 ºC for more than 15 minutes is recommended by EN13060 or EN ISO17665-1.
CAUTION
• Do not autoclave the Control Unit and the Foot Control.
• Never sterilize the Handpiece, etc., with ultraviolet radiation, which might cause discoloration.
• If you autoclave the Handpiece together with any instrument to which antiseptic solution adheres, discoloration
might occur.
• If any saline solution or stain adheres to the Control Unit, well wipe it off with a cloth impregnated with clean water
and squeezed out. Then, wipe the Control Unit well with a dry cloth.
• Do not clean the Control Unit with any solvent such as thinner, benzene, etc.
• We do not recommend you to sterilize on plasma sterilization or EOG sterilization.
8-2-4 Cleaning of Optic Fiber (Optic Handpiece)
Wipe the debris off the end of the Optic Fibers at the Handpiece
with alcohol soaked cotton swab (Fig. 27).
CAUTION
Do not use any sharp pointed tools to clean the Optic Fiber End
Face. In case the light degradation, contact dealer.

10. Troubleshooting

When you suspect a failure, refer to the following before requesting a repair. If your problem does not fall into any of the
following, or troubleshooting as suggested does not improve the problem, contact dealer as it is likely that the product is
faulty.
Case
Control Unit cannot be turned ON
(LCD display does not turn on).
Possible Cause
AC Power Cord is not connected to Control
Unit.
AC Power Cord is not plugged into wall
outlet.
Fuse has blown.
18
Fiber end-face
Remedy
Make sure it is connected.
Make sure it is plugged in.
Contact dealer. *
Cotton swab
Fig.27