Autoclave sterilization is recommended.
Autoclave sterilization is required for the first time use and after each patient as noted below.
The following items can be autoclaved.
Handpiece with Handpiece Cord (Optic/Non-Optic), Tip, Tip Wrench, Tips Holder, Handpiece stand, Tube Holder, and
Sterilization Cassette.
8-2-1 Handpiece
Optic Handpiece
1) Disconnect the Handpiece cord from the Control Unit.
2) Remove the Irrigation Tube from the Handpiece and dispose of it.
3) Brush (do not use a metal brush) dirt off the surface, and then wipe with alcohol soaked cloth.
4) Set the Handpiece with Handpiece Cord in the Sterilization Cassette.
Non-Optic Handpiece
1) Turn the Nose Cover counter-clockwise to remove.
2) Brush any contaminants from inside the Nose Area with a soft
brush. (Do not use metal brush) Wipe with alcohol soaked cloth
(Fig. 26).
3) Replace the Nose Cover onto the Handpiece turn clockwise.
4) Set the Handpiece with Handpiece Cord in the Sterilization
Cassette.
CAUTION
Only handpiece with cord can be washed via Thermo Disinfector.
* Confirmed Thermo Disinfector is "Type: G7882" (Miele & Cie)
8-2-2 Tip, Tip Wrench, and Tips Holder
1) Brush (do not use a metal brush) dirt off the surfaces of the Tip, Tip Wrench, and Tips Holder, and then wipe with
alcohol soaked cloth.
2) Mount the Tip to the Tips Holder before setting it in the Sterilization Cassette. The case accommodates up to 2
Tips Holders.
3) Set the Tip Wrench in the Sterilization Cassette.
CAUTION
Use rubbing alcohol only.
* A Sterilization Cassette with a Handpiece and other items set in it.
17
Fig.26