NewLife Intensity 10
EMC Testing
Medical equipment needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in this section.
Guidance and Manufacturer's Declaration —Electromagnetic Emissions
The NewLife is intended for use in the electromagnetic environment specified below. The customer or the user of
the NewLife should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
22 - ENG PN MN239-C4 F | User Manual
Compliance
Electromagnetic Environment - Guidance
The NewLife uses RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Class B
The NewLife is suitable for use in all establishments, including do-
mestic establishments and those directly connected to the public
Complies
low-voltage power supply network that supplies buildings used for
domestic purposes.
Complies