Figura C-16:
Menú Herramientas de servicio: Mantenimiento
1
Routine Maintenance
1 Trim mA Output 1
2 Trim mA Output 2
3 Meter Verification *
3
Meter Verification **
1 Run Meter Verification
2 View Test Results
3 Schedule Meter
Verification
*
Displayed only if meter verification is
enabled.
** Displayed only if Smart Meter
Verification is enabled. For earlier
versions, the Meter Verification
Method is launched.
Manual de configuración y uso
On-Line Menu >
3 Service Tools >
4 Maintenance
2
Zero Calibration
1 Mass Flow Rate
2 Volume Flow Rate
3 Zero Time
4 Zero Value
5 Standard Deviation
6 Perform Auto Zero
7 Restore Factory Zero
4
Temperature Calibration
1 Temperature
2 Temp Cal Factor
3 Offset Calibration
4 Slope Calibration
Uso del Comunicador de Campo con el transmisor
3
Density Calibration
1 Density
2 Dens Pt1 (Air)
3 Dens Pt2 (Water)
4 Dens Pt3 T-Series
5 Dens Pt4 T-Series
6 Flowing Dens (FD)
5
Diagnostic Variables
1 Sensor Model
2 Drive Gain
3 Input Voltage
4 LPO Amplitude
5 RPO Amplitude
6 Board Temperature
7 Tube Frequency
8 Live Zero
9 Fld Verification Zero