Example: 100 µl Air-displacement pipette, dynamic test
Before testing, mount a new, unused, disposable tip, and insert the pipette with the tip in the previously
selected and attached Adapter 1 (see page 46).
What to do?
1.
Calling up the
Main Menu
2.
Selecting the
pipette type
3.
Confirming the
pipette type
4.
Selecting the
single-channel
pipette
5.
Confirming the
single-channel
pipette
6.
Selecting the
pipette volume
Testing single-channel pipettes
How to do it
The 'Main Menu' is called
up by pressing or turning the
select knob.
Set to 'Pipette Type' by
turning the select knob.
Press the 'Execute' button.
The 'Channels' submenu is
called up.
Set to 'Single-Channel' by
turning the select knob.
Press the 'Execute' button.
The 'Nominal Volume'
submenu is called up.
'100 µl' is selected by turn-
ing the select knob.
Continued on next page
Keys to press
Display readout
Pipette Type
Self Test
Switch OFF
Settings...
select
Pipette Type
Self Test
Switch OFF
Settings...
select
Single Channel
Multichannel
select
Single Channel
Multichannel
select
Nominal Volume
25 µl
50 µl
100 µl
200 µl
select
Nominal Volume
25 µl
50 µl
100 µl
200 µl
select
Main Menu
back
Main Menu
back
Channels
back
Channels
back
back
back
47