Troubleshooting
Error during the self-test
Self test
error: 1XXX
see instruction manual
restart
Error number
1000
1010
1020
1040
1050
1030
1060
1070
1080
1090
1020
1150
1160
1170
1200
64
Error display
back
Cause
Pressure sensor is defective Send the instrument out for repair.
Pump running continuously
Pumping down takes too
long
Defective valve
Aeration through the aera-
tion channel takes too long
Instrument leaks
Aeration through the
adapter takes too long
Electronic defect
Corrective action
Send the instrument out for repair.
Check whether the caps are firmly attached, and
that the adapter is firmly screwed in.
If they already are, then the pump is defective.
Send the instrument out for repair.
Send the instrument out for repair.
Send the instrument out for repair.
Check the caps, adapter seating, and adapter filter,
and replace if necessary. Repeat the self-test. If
the error is not resolved, send the instrument out
for repair.
Remove the cap if this hasn't been done. If the
adapter filter is contaminated, replace the filter.
Send the instrument out for repair.