Testing multichannel pipettes
Example: 200 µl 8-channel air-displacement pipette, dynamic test
Before testing, mount a new, unused, disposable tip, and insert four channels with tips in the previously
selected and attached Adapter 3 (see page 46).
What to do?
1.
Calling up the
Main Menu
2.
Selecting the
pipette type
3.
Confirming the
pipette type
4.
Selecting the
multichannel
pipette
5.
Confirming the
multichannel
pipette
6.
Selecting the
pipette volume
50
How to do it
The 'Main Menu' is called
up by pressing or turning the
select knob.
Set to 'Pipette Type' by
turning the select knob.
Press the 'Execute' button.
The 'Channels' submenu is
called up.
Set to 'Multichannel' by
turning the select knob.
Press the 'Execute' button.
The 'Nominal Volume'
submenu is called up.
'200 µl' is selected by turn-
ing the select knob.
Keys to press
Display readout
Pipette Type
Self Test
Switch OFF
Settings...
select
Pipette Type
Self Test
Switch OFF
Settings...
select
Single Channel
Multichannel
select
Single Channel
Multichannel
select
Nominal Volume
50 µl
100 µl
200 µl
250 µl
select
Nominal Volume
50 µl
100 µl
200 µl
250 µl
select
Main Menu
back
Main Menu
back
Channels
back
Channels
back
back
back