Performing a Control Solution Test

1.
Check control solution expiration date.
2.
Remove test strip from vial.
3.
Insert test strip to turn meter on.
Note: The meter turns off after 2 minutes of inactivity.
Remove and reinsert the unused test strip to restart
the meter.
If the meter does not turn on go to Troubleshooting section.
IMPORTANT: Do not use control solution past expiration date. Discard
control solution 3 months after opening or on the expiration date
printed on the bottle whichever comes first. (Example: open April 15,
discard July 15; write the discard date on the side of the bottle.)
Top
Printed side up
Sample areas