Serial Interface - Hameg Instruments HM8115-2 Manual De Instrucciones

Ocultar thumbs Ver también para HM8115-2:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

S e r i a l i n t e r f a c e

Serial Interface

The HM8115-2 is well equipped for use in automated test sy-
stems. An optcoupler-isolated RS-232 interface is standard.
Interface parameters
N, 8, 1, Xon-Xoff:
(No parity bit, 8 data bits, 1 stop bit, Xon-Xoff.
A terminal program like HyperTerminal may be used for da-
ta transmission. After performing all settings in the terminal
program press the ENTER key once prior to sending the fi rst
command to the HM8115-2
Baud rate
1200 or 9600 baud.
Changing interface parameters
Only the baud rate can be selected as either 1200 or 9600 baud.
In order to do this proceed as follows:
Turn off the HM8115-2.
Turn the instrument back on.
Press the lefthand FUNCTION pushbutton
Press the lefthand pushbutton
until the LED "WATT" lights up.
This new baud rate will be stored permanently unless
changed.
Serial interface
15
The RS-232 interface connector is located on the rear panel
(9pin submin D). The interface allows the transmission of data
from the instrument and its remote control.
F o r t h e c o n n e c t i o n b et w e e n t h e HM 8115 -2 a n d a
PC (COM port) any standard cable with 9pin submin D on both
sides may be used, provided it is shielded and < 3 m.
If a PC has a 25pin connector an adapter 25 to 9pin
has to be inserted, only 3 wires are used.
HINT
Connections:
POWER METER
Pin
name / function
2
Tx Data / output
3
Rx Data / input
5
Ground
38
Subject to change without notice
and keep it depressed
16
17
PC COM Port (9poles)
Pin
name / function
2
Rx Data / input
3
Tx Data / output
5
Ground
Software
1. Installation
For the installation of the software HM8115-2 please start the
fi le setup.exe and follow the instructions of the installation
assistant.
2. The program
The software HM8115-2 was developed for the programmable
Hameg Instruments Power Meter HM8115-2. The power meter
can be connected with the PC by 3 different interfaces: Serial
interface (standard), USB interface (HO870) and GPIB interface
(HO880). If the USB interface is used, a virtual COM port has
to be installed. The COM port driver can be downloaded from
our website www.hameg.com
Picture 2.1: User interface
IMPORTANT WITH GPIB!
This software only supports National Instruments
GPIB cards and cards, that are fully compatible
with the National Instruments cards.
STOP
The program is devided into 4 parts (Settings, Control panel,
Measurement, Instruction), which are described in the follow-
ing:
2.1 Settings
Six different parameters can be set:
Interface:
In this fi eld the interface can be selected,
which is used for the connection to the PC.
Available settings: Com1-4, GPIB
GPIB address:
Setting of the GPIB address the HM8115-
2(only with GPIB)
Beep enable:
Activation/Deactivation of acoustic signals
Show data traffi c: This option offers the possibility of letting the
data traffi c in the editing fi elds 'Command'
and 'Answer' indicate.
Device locked:
With this option activated the instrument on-
ly can be operated by software. The control
elements are locked!

Publicidad

Tabla de contenido
loading

Tabla de contenido