Emergency; Emergency Stop - Vermeiren Albatros II Manual De Instrucciones

Ocultar thumbs Ver también para Albatros II:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 118
2.6

Emergency

In cases for emergency the emergency stop and the function for lowering in an emergency
must be used.

2.6.1 Emergency stop

2.6.2 Manual emergency lowering
L
WARNING: Risk of injury – Adjust the manual emergency lowering for the weight
of the patient.
1
Please note that the manual emergency lowering is only possible when the patient is sitting
in the patient lift.
Pull force: 10 mm for force from 80 à 100N.
In emergencies, and during assembly and disassembly,
the emergency stop is activated by pressing the red
button on the control box.
Push on the button and turn in the direction of the arrows.
If the button is fully pressed to the back turn back in the
opposite direction of the arrows.
The emergency stop is deactivated by
turning the red button in the direction
of the arrows.
Emergency lowering if the power fails or the batteries are
empty can be done using the red pull button (1) at the
lower end of the motor. This emergency lowering is
factory-set for a patient weight of 50 kg.
Standard calibration: 2500N, 15 mm/s to lower the
patiënt.