Replacing The Fuses - Surtek 111001 Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

E N G L I S H
5.
Replace
the
battery with a new
9V battery (NEDA
1604 or 6F22 or
006P).
6. Rejoin the case
bottom and case
top, and reinstall
the screw.
C.

Replacing the Fuses

To avoid electrical shock or arc blast, or
personal injury or damage to the Meter,
use specified fuses ONLY in accordance
with the following procedure.
42
Figure 7
(Figure 7)
Warning
To replace the Meter's fuse:
1. Disconnect the connection between the testing
leads and the circuit under test, and remove the
testing leads away from the input terminals of the
Meter.
2. Turn the Meter to OFF position.
3. Remove the screw from case bottom, and
separate he case bottom from the case top.
4. Remove the fuse by gently prying one end
loose, and then take out the fuse from its bracket.
5. Install ONLY replacement fuses with the
identical type and specification as follows and
make sure the fuse is fixed firmly in the bracket.
0,3A, 250V, fast type, φ5x20 mm.
6. Rejoin the case bottom and case top, and
reinstall the screw.
Replacement of the fuses is seldom required. Burning of
a fuse always results from improper operation.

Hide quick links:

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido