The surgical aspirator SUPER TOBI is intended for use in the electromagnetic environment specified below.
The customers or the user of the surgical aspirator SUPER TOBI should assure that it's used in such an environment.
Immunity Test
Conducted Immunity
3Vrms 150kHz to 80Mhz
IEC / EN 61000-4-6
(for non life-supporting
Radiated Immunity
3V/m 80MHz to 2.5GHz
IEC / EN 61000-4-3
(for non life-supporting
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the absorption and by reflection
from buildings, objects and people.
a) The field intensity for fixed transmitters such as the base stations for radiotelephones (mobile and cordless) and terrestrial mobile radio, amateur
radio devices, radio AM and FM transmitters and TV transmitters can not be theoretically and accurately foreseen.
To establish an electro-magnetic environment generated by fixed RF transmitters, an electro-magnetic study of the site should be considered. If the
field intensity measured in the place where the device will be used surpasses the above mentioned applicable level of conformity, the normal
functioning of the device should be monitored. If abnormal performance arises, additional measures such as changing the device's direction or
positioning may be necessary.
b) The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 3 V/m.
Recommended separation distance between portable and mobile radio-communication devices and the monitor
The SUPER TOBI surgical aspirator is intended to operate in an electro-magnetic environment where RF irradiated interferences are under control. The
client or operator of the SUPER TOBI device can help prevent electro-magnetic interference by keeping a minimum distance between the portable and
mobile RF communication devices (transmitters) and the SUPER TOBI device, as recommended below, in relation to the radio-communication
maximum output power.
Maximum nominal output
power of the
Transmitter W
0,01
0,1
1
10
100
For transmitters with a maximum nominal output power not shown above, the recommended separation distance in metres (m) can be calculated using
the equation applicable to the transmitter frequency, where P is the maximum nominal output power of the transmitter in Watt (W) depending on the
transmitter's manufacturer.
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the absorption and by the reflection
from buildings, objects and people.
Guidance and manufacturer's declaration – Immunity Emissions
Immunity Test
Immunity Test
devices)
devices)
Separation distance from the frequency transmitter (m)
150 kHz to 80 MHz
P
d = [3.5 / V
]
1
0,12
0,38
1,2
3,8
12
V
= 3 V rms
The portable and mobile RF communication devices, including
1
cables, must not be used closer to the SUPER TOBI device, than
the separation distance calculated by the equation applicable to
the transmitter frequency. Recommended separation distance
d = [3.5 / V
d = [12 / E
d = [23 / E
Where P is the maximum nominal output voltage of the
transmitter in Watt (W) depending on the manufacturer of the
transmitter and the recommended separation distance in metres
(m). The intensity of the field from the fixed RF transmitters, as
determined by an electro-magnetic study of the site
lower than the level of conformity of each frequency interval
It is possible to check for interference in proximity to devices
identified by the following symbol:
80 MHz to 800 MHz
P
d = [12/E
]
1
0,12
0,38
1,2
3,8
12
12
Immunity Test
P
]
1
P
]
from 80 MHz to 800MHz
1
P
]
from 800 MHz to 2.5 GHz
1
800 MHz to 2.5 GHz
d = [23/E
]
1
0,23
0,73
2,3
7,3
23
a)
, could be
.
b)
P

Publicidad