4. Push TOC CALIBRATION.
5. Select default or custom settings.
6. Obey the screen prompts to set the number of repetitions and confirm the sucrose standard
values.
Note: To save the new settings as the default, push SAVE AS DEFAULT.
7. Install the standards in the order shown on the screen. If the bottle bay door closes too soon,
push UNLOCK to open the door.
8. Push START.
9. When the calibration is complete, the results are shown. If a printer is attached, the report can be
sent to a printer. Select the necessary action to complete the calibration.
• Push CHECK to accept the calibration and to remove the bottles. This option will not show if
the Autoaccept setting is enabled.
• Push CANCEL to repeat the calibration and discard the results.
• Push EXPORT to send the data to the supplied USB flash drive.
Conductivity calibration
Use the optional calibration resistor for this procedure.
resistor.
1. Make sure that the conductivity standards are at room temperature.
2. Push << to show the sliding menu.
3. Push BOTTLE MODE, then RUN STANDARDS.
4. Push CONDUCTIVITY CALIBRATION.
5. Obey the screen prompts.
6. Install the standards in the order shown on the screen. If the bottle bay door closes too soon,
push UNLOCK to open the door.
7. Push START. Install the resistor when the screen prompt shows.
Note: When the calibration is complete, the results show on the screen. If a printer is attached, the report prints
automatically.
8. Select the necessary action to complete the calibration.
• Push CHECK to accept the calibration and to remove the bottles. This option will not show if
the Autoaccept setting is enabled.
• Push CANCEL to repeat the calibration and discard the results.
• Push EXPORT to send the data to the supplied USB flash drive.
18 English
Figure 9
shows the location of the calibration