Servicing Dentsply Sirona Quick Coupling; Replacing O-Rings - Dentsply Sirona Midwest Stylus Plus Instrucciones De Uso

Ocultar thumbs Ver también para Midwest Stylus Plus:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6607290.book Seite 59 Donnerstag, 4. Januar 2018 11:55 11
®
Operating Instructions Midwest Stylus
Plus / Midwest Tradition
A
B
66 07 290 D3673.201.01.04.02
®
TC/Pro TCF
Tip: Record the time and result of the check for your own information.
8.2

Servicing Dentsply Sirona quick coupling

8.2.1

Replacing O-rings

1. Remove the defective O-ring.
Using T1 Spray
2. Slide the tool (A), with the new O-ring attached, to just in front of the
corresponding groove (B).
3. Insert the O-ring. Do not use any sharp objects.
4. Lightly oil the O-ring with Midwest Plus Aerosol Spray.
NOTICE! Do not use Vaseline or silicone grease on the O-rings.
8 Maintenance
59

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Midwest tradition tcMidwest tradition pro tcf

Tabla de contenido