SIBELSOUND
If there is an error during the export process, a warning message appears indicating the type of error.
For more information about export files format, please contact
AUTOMATIC AUDIOMETRY (HUGHSON-WESTLAKE) OPTION
The AUTOMATIC AUDIOMETRY (HUGHSON-WESTLAKE) option allows users to perform an air-conduction tonal test (without masking)
in a fast, simple manner, with minimum technician intervention.
To personalize various parameters for the automatic test, access the TEST CONFIGURATION menu
appears, where the following automatic test items can be customized:
•
Threshold method (2/3 or 3/5)
•
Tone type (Continuous or Pulsating)
•
Starting level (between 0 and 70 dB)
•
Random Off time (between 0 and 2 seconds)
•
Tone duration (between 0.5 and 2 seconds)
•
Frequencies to examine (125, 250, 500, 750, 1500, 2000, 3000, 4000, 6000, 8000 Hz). The 1000 Hz frequency is always
examined, and does not appear on this screen.
520-800-MUM · Rev 1.01
DUO
· User's manual
SIBEL
S.A.U.Customer Service.
EN 26
and press
. A screen