AUTOMATIC AUDIOMETRY (HUGHSON-WESTLAKE) OPTION
The AUTOMATIC AUDIOMETRY (HUGHSON-WESTLAKE) option allows users to perform an air-conduction tonal test (without masking)
in a fast, simple manner, with minimum technician intervention.
To personalize various parameters for the automatic test, access the TEST CONFIGURATION menu
appears, where the following automatic test items can be customized:
Threshold method (2/3 or 3/5)
Tone type (Continuous or Pulsating)
Starting level (between 0 and 70 dB)
Random Off time (between 0 and 2 seconds)
Tone duration (between 0.5 and 2 seconds)
Frequencies to examine (125, 250, 500, 750, 1500, 2000, 3000, 4000, 6000, 8000 Hz). The 1000 Hz frequency is always
examined, and does not appear on this screen.