Dräger X-am 5600 Instrucciones De Uso página 13

Ocultar thumbs Ver también para X-am 5600:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 32
Areas subject to explosion hazards, classified by divisions.
The instrument is intended for use in explosion-hazard areas, in which gas
classified by class I&II, Div. 1 or Div. 2 may occur.
It is intended for use in a temperature range
when using power pack ABT 0100:
o
o
o
of –20
C to +50
C or –20
C to +40 °C depending on the batteries used
when using power pack HBT 0000/HBT 0100:
o
from –20
C to +50 °C
and for areas in which gases or dusts of groups A, B, C, D or E, F, G, and
temperature class T3 or T4 (depending on the batteries and rechargeable
battery) may be present.
Safety Instructions
To reduce the risk of ignition of a flammable or explosive atmosphere, strictly
adhere to the following Caution and Warning statements:
Only use power packs ABT 0100 (83 22 237), HBT 0000 (83 18 704) or
HBT 0100 (83 22 244). See marking on power pack for approved batteries and
related temperature class.
Do not change batteries in areas where there is a danger of explosion.
To reduce the danger of explosion, do not mix new batteries with old batteries
and do not mix batteries made by different manufacturers.
Always disconnect the instrument from the power pack before carrying out any
maintenance operations.
Substitution of components may impair intrinsic safety.
Not tested in oxygen enriched atmospheres (>21% O
WARNING
WARNING
WARNING
WARNING
WARNING
CAUTION
).
2
What is what
1
2
2
3
8
+H
2
X-am 5600
7
6
4
0
5
1
Gas entry
2
Alarm LED
3
Buzzer
4
[ OK ] key
5
Power pack
6
[ + ] key
7
Display
8
Display labels
H
added signal (optional)
2
Special symbols:
Fault message
Warning message
Display peak value
Show TWA
Show STEL
Bump test mode
Fresh air calibration
CO
2
Vol%
9
ch
4
%LEL
10
O
2
Vol%
CO
ppm
H
S
2
ppm
11
NH
3
ppm
12
13
14
15
9
IR interface
10
Fastening clip
11
Type plate
12
Charging contacts
13
Measured gas display
14
Measured value display
15
Special symbols
1-button calibration
Single gas calibration
Password required
Battery 100 % full
Battery 2/3 full
Battery 1/3 full
Battery empty
13

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Mqg 01 serieMqg 010

Tabla de contenido