Replacing the batteries / rechargeable batteries

Danger of explosion!
Do not throw used batteries into fire or try to open them by force.
Dispose of the batteries in accordance with national regulations.
Danger of explosion!
Do not replace the batteries / rechargeable batteries in areas where there is a
danger of explosion.
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
Alkaline batteries – T3 – (not rechargeable!)
Panasonic LR6 Powerline
Varta Type 4106 (power one) oder
Varta Type 4006 (industrial)
Alkaline batteries – T4 – (not rechargeable!)
Duracell Procell MN1500
NiMH rechargeable batteries – T3 – (rechargeable)
GP 180AAHC (1800) max. 40
Charge the type HBT 0000 or HBT 0100 power pack using the relevant Dräger
charger. Charge NiMH rechargeable batteries for battery holder ABT 0100 in
accordance with the manufacturer's specifications. Ambient temperature
during the charge process: 0 to +40 °C.
Switching off the instrument:
z
Press and hold the [ OK ] key and the [ + ] key simultaneously.
z
Loosen the screw on the power pack and remove the power pack.
For the battery holder (order no. 83 22 237):
Danger of explosion!
The Dräger X-am 5600 must only be operated with battery holder ABT 0100
(X-am 5600), identified with a silver sticker.
z
Replace alkaline batteries or NiMH rechargeable batteries. Ensure correct
polarity.
WARNING
WARNING
o
C ambient temperature.
WARNING
WARNING
For the NiMH power pack T4 (order no. 83 18 704) / T4 (order no. 83 22 244):
z
Completely replace the power pack.
z
Insert the power pack into the instrument and tighten the screw, the
instrument switches on automatically.
Charge the instrument with NiMH power pack T4
(83 18 704) / T4 HC (83 22 244)
Danger of explosion!
Do not charge underground or in explosion-hazard areas!
The chargers are not designed in accordance with the regulations for firedamp
and explosion protection.
Charge the type HBT 0000 or HBT 0100 power pack using the relevant Dräger
charger. Ambient temperature during the charge process: 0 to +40 °C.
Even if the instrument is not used, we recommend that you store the instrument
in the charger!
z
Insert the switched off instrument into the charger.
Display LED on the charger:
To maintain the lifetime of the batteries, charging is temperature controlled and
only performed in a temperature range of 5 to 35
range is left, the charging is automatically interrupted and automatically
continued after the temperature range has been reached again. The charging
time is typically 4 hours. A new NiMH power pack reaches its full capacity after
three complete load/unload cycles. Never store the instrument for extended
periods (maximum 2 months) without being connected to a power source, as
this drains the internal buffer battery.
WARNING
WARNING
Charge
Fault
Full
o
C. When this temperature
17