Guidance and manufacturer's declaration – electromagnetic immunity
The Elipar DeepCure-L is intended for use in the electromagnetic environment specified below.
The customer or the user of the Elipar DeepCure-L should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrostatic
± 2 kV for power
transient / burst
supply lines
IEC 61000-4-4
± 1 kV for input/output
lines
Surge
± 1 kV differential
mode
IEC 61000-4-5
± 2 kV common mode
Voltage dips, short
< 5 % U
interruptions and
(> 95 % dip in U
voltage variations
for 0,5 cycle
on power supply
input lines
40 % U
(60 % dip in U
IEC 61000-4-11
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
< 5 % U
(>95 % dip in U
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE
U
126
T
)
T
T
)
T
T
)
T
T
)
T
is the a. c. mains voltage prior to application of the test level.
T
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 1 kV differential
mode
± 2 kV common mode
< 5 % U
T
(> 95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
< 5 % U
T
(> 95 % dip in U
)
T
for 5 sec
3 A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Elipar DeepCure-L requires
continued operation during power
mains interruptions, it is recommended
that the Elipar DeepCure-L be powered
froman uninterruptible power supply or
a battery.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.