the arrow keys to change the value of the calibration weight. When it will be correct press the key
•
The display will show "load". Place the calibration weight on the platform. Press the key
•
If the calibration is correct the display will return to normal mode, if an error message appears, try to recalibrate, because perhaps some
factor could impede proper calibration.
•
If the problem persists please contact Baxtran or its supplier. After calibration, the viewer should be reviewed to verify that the calibration and
linearity are correct. If necessary repeat calibration, especially make sure that the viewer is in a stable position before accepting any weight.
11. ERROR CODES
ERROR CODES
- - - - -
Err 1
Err 2
Err 4
Err 6
12. TECHNICAL CONFIGURATION
C
Press the key
when the viewer will be in normal weight mode and the display will show"
Press the key
until the display shows "
The display must show "
parameter.
Press the key
to exit.
44
DESCRIPTION
Above the range
Error when configuring the
date
Error when configuring time
Configuration zero error
A/D converter out of range
PROG
" and press
PIN
", then press the keys
DETERMINATION
Remove the weight of the platform. If the problem persists contact
BAXTRAN or with your distributor for assistance.
Introduce the date using the correct format and the adequate values.
Formato: aa.mm.dd.
Introduce the date using the correct format and the adequate values.
Formato: hh.mm.ss.
The indicator was not correctly configured in the 0 range or it was
turned on while the
key was pressured. Remove the weright
from the platform and press again.
Us the key
for set the display in 0 value.
If the problem persists please contact Baxtran or its supplier.
The values of the A/D converter are outside the normal range. Remove
the weight of the platform if there is an excess of weight. Indicates that
the load cell or electronics may not be functioning properly.
for enter.
,
and
to enter the configuration mode and the key
USER'S MANUAL BR40
.
.
".
F0 H-L
To select the
EN