Table 2
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electrostatic
±6 kV cont act
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
power supply lines:
transient/burst
±2 kV
IEC 61000-4-4
input/output lines:
±1 kV
Surge
line(s) to line(s):
IEC61000-4-5
±1 kV
line(s) to earth:
±2 kV
100 kHz repetition
frequency
Voltage
<5% UT (>95%
dips, short
dip in UT ) for 0.5
interruptions
cycle
and voltage
40% UT (60% dip in
variations on
UT ) for 5 cycles
power supply
70% UT (30% dip in
input lines
UT ) for 25 cycles
IEC 61000-4-11
<5% UT (>95% dip
in UT ) for 5 s
Power
3 A/m
frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
±6 kV cont act
±8 kV air
not applicable
not applicable
not applicable
3 A/m
EMC GUIDANCE
Electromagnetic
environment - guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should
be that of a typical commercial
or hospital environment. If
the user of the device requires
continued operation during
power mains interruptions,
it is recommended that the
device be powered from an
uninterruptible power supply
or a battery.
Power frequency magnetic
fields should be at levels cha-
racteristic of a typical location
in a typical commercial or
hospital environment.
145