Prosthesis settings should be optimised using the adjustment software.
The remote control is not intended for use by the prosthetist to set up the
prosthesis. The behaviour of the prosthesis can be changed to a certain
extent on a day-to-day basis using the remote control (e.g. while becom
ing accustomed to the prosthesis). The prosthetist can use the adjust
ment software to trace these changes at the next appointment.
•
If the settings of a MyMode are to be modified, one must first switch to
this MyMode.
8.2.1 Using remote control to change prosthesis settings
INFORMATION
The number and names of the configurable parameters depend on the
respective prosthesis. Therefore, the illustrations of menus in this section
are by way of example only.
Further information on the parameters is found in the instructions for use of
the respective prosthesis.
12
1) Use the
,
main menu and confirm this with the key.
→ A list appears with the parameters for the currently active mode
(see fig. 13).
2) Use the
,
with the key.
→ A scale will appear in the display of the remote control and shows
the value currently set for the parameter as well as the parameter's
limit values (see fig. 14).
3) Use the , keys to change the parameter setting and confirm this with
the key.
To leave the menu without saving changes, use the
the
symbol and confirm this with the key.
→ The message "Successful" appears briefly to confirm these
changes were saved in the prosthesis.
4) If necessary, switch off the remote control unit by selecting and confirm
ing the menu option "RC power off".
32 | Ottobock
13
keys to select the menu option "Device settings" in the
keys to select the parameter you want and confirm this
14
,
keys to select