Error message
Device disconnec
ted
No new device
found
Cause
A current connec
tion to the pros
thesis was inter
rupted.
No new prosthesis
could be found
since this pros
thesis is already
stored.
Correction
3) See section "Managing pros
theses" (see Page 35) for fur
ther information.
Check the following points:
•
Distance from the prosthesis
to the remote control
•
Was the prosthesis connec
ted to another remote con
trol?
The
connection
already saved in the remote
control must be deleted in
order to re-connect the pros
thesis to this remote control.
A new connection must then
be established from the pros
thesis to this remote control
(see Page 35).
•
Use the
key to confirm the
error message and try to
establish
again.
1) Confirm error message with
the key.
2) Use the
,
the
menu
device" and confirm this with
the key.
3) Use the
,
the prosthesis required and
confirm this with the key.
that
was
the
connection
keys to select
option
"Select
keys to select
Ottobock | 43