TROUBLESHOOTING GUIDE

Problem
On/off button
flashing
notification
On/off button isn't
responding
Condensation
Unit is leaking
from blue
connectors
If the problem continues, please contact Breg Customer Care at 1-800-321-0607
or +1-760-795-5440 for assistance.
18
Possible Solutions
• Restart the Unit
○ Turn unit off by either pressing the on/off button or by disconnecting
and reconnecting power supply from the unit or from the wall.
○ Press on/off button to turn unit back on.
○ Press cold or compression therapy buttons (if required) to start desired
therapies.
• If compression therapy does not function after pressing compression button
(cold-compression pad doesn't appear to be inflating or compression
therapy setting light is not on), see Compression Therapy troubleshooting
sections on page 19.
• If cold therapy does not function after pressing cold button, see Cold
Therapy troubleshooting section on page 19.
• Confirm power outlet is working and power supply is fully inserted in the
outlet.
• Confirm that power supply plug is fully inserted into the unit lid.
• Confirm the correct Polar Care Wave power supply is connected (see page 20).
• Wrap material loosely over cold-compression pad and tubing to minimize
air exposure.
• Protect the wound site by using a sterile dressing with waterproof barrier.
Acceptable barriers include Webril
Note: Some condensation on the tube and cold-compression pad is possible,
especially in warmer climates. Place a cloth layer under the container to protect
items under the unit.
• Disconnect blue tabs on the cold-compression pad connector.
• Check that the Breg logos on the blue connectors are lined up correctly and
reconnect.
, Kerlix
, cast padding and elastic bandage.