TROUBLESHOOTING GUIDE
Problem
Cold therapy:
water not
flowing to
cold-compression
pad or pad not
cold
Compression
therapy:
cold-compression
pad not
compressing or
filling with air
Compression
therapy:
cold-compression
pad not deflating
properly
If the problem continues, please contact Breg Customer Care at 1-800-321-0607
or +1-760-795-5440 for assistance.
Possible Solutions
• Confirm on/off button is illuminated and a cold therapy "colder" or
"cold" setting light is on to indicate that cold therapy is functioning.
• Check that blue connectors between cold-compression pad and tubing are
fully engaged with Breg logos lined up. Disconnect blue connectors and
reconnect if needed.
• Allow 10 minutes for cold-compression pad temperature to stabilize.
• Confirm unit has both ice and water to appropriate levels.
• While using the Breg Ice Bags, cubed ice may also be added to the
container up to the ice fill line to optimize the colder setting.
• Elevating the unit to the same or higher elevation than the cold-compression
pad may increase water flow to the cold-compression pad.
• Check that tubing and cold-compression pad are not kinked.
• Remove cold-compression pad and lay it flat. Allow cold-compression pad to
fill with water and then reapply.
Note: During compression therapy operation, it is normal for the compression to
inflate and deflate continuously. The duration of each inflation-deflation cycle is
approximately 1 to 2 minutes.
• Confirm on/off button is illuminated and a compression therapy "regular" or
"low" setting light is on to indicate that compression therapy is functioning.
• Check that blue connectors between cold-compression pad and tubing are
fully engaged with Breg logos lined up. Disconnect blue connectors and
reconnect if needed.
• Tighten cold-compression pad.
• Check that tubing and cold-compression pad are not kinked.
• Remove cold-compression pad and lay it flat. Check if the cold-compression
pad is inflating with air and then reapply.
• Check that blue connectors between cold-compression pad and tubing are
fully engaged with Breg logos lined up.
• Disconnect blue connectors, deflate pad by rolling to remove air, and
reconnecting blue connectors if needed.
19