-Sterilisation
Sterilisation in an autoclave conforming to: EN 13060,
EN ISO 17665-1
CAUTION!
- Follow the instructions of the autoclave manufacturer.
- Clean, disinfect and lubricate the handpiece before
sterilisation. The presence of blood residues on the external
or internal surfaces can lead to the formation of corrosion.
- Make sure that the handpiece is properly dried after
sterilisation.
Authorised sterilisation procedures
1) Seal the handpiece and its accessories in sterilisation
bags conforming to EN 868-5.
2) Sterilise the bagged handpiece in an autoclave
according to EN ISO 17665-1 standard and respect the
following conditions:
- Class B steam sterilisation in autoclaves conforming to EN
13060 standard. Sterilisation time at least 3.5 minutes at 135 °C.
3) Remove the bag from the autoclave.
-Storage
- Keep the handpiece in the sterilisation bag until it is needed.
- Once sterilised, store the product in a dry and dust-free
place, if possible with a low microbial charge.
CAUTION!
Respect the validity date for the sterilisation!
-replacing the O-rings
Replace the O-rings if you notice any water leaks betwe-
en the micromotor and the handpiece.
1) Squeeze the old O-ring between your fingers to remove it.
2) Fit the new O-ring in the correct seat.