SPORT-ELEC Multisportpro Manual De Uso página 10

Ocultar thumbs Ver también para Multisportpro:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 16
END OF PROGRAM
Once the program has finished, switch off the device and unplug it from the belt.
Remove the electrodes and moisten them slightly with a few drops of water before replacing them onto their transparent plastic
backing sheets.
IV. REcOMMEnDaTIOnS OF USE
The unit is recommended for physiotherapists for functional re-education.
You can restart a program's cycle on another muscle zone or change the program while remaining on the same muscle zone,
f
f
without exceeding however 40 min use per muscle zone and per day.
Clean the unit and its cables with a cloth dampened with a disinfecting solution (Mercryl type) diluted in water at 1/10
f
f
The adhesive electrodes can be rehydrated before and after use with a disinfecting agent (e.g. Hexamedine).
f
f
The adhesive electrodes must be repositioned on their support, and stored in a fresh storage between +5°C and +10°C.
f
f
The adhesive electrodes have a lifetime of about 40 uses, you can order them directly from the online shop www.sport-elec.com,
f
f
or from your reseller.
PRECAUTIONS OF USE :
Do not use the unit on your face
f
f
Do not leave within reach of children
f
f
Ask for advice from your physician :
if you have electronic or metallic implants (pins, etc.)
f
f
in case of skin condition (wounds,...)
f
f
TROUbLE SHOOTING
If your device doesn't work properly, do not use it and contact
your retailer.
RECOMMENDATIONS :
Please apply regulations enforced in your country
when disposing your product at its end of life.
Therefore, please discard it in a location planned to
this effect, ensuring it is disposed in an environment-
friendly manner.
CONTRA-INDICATIONS :
Do not use SPORT-ELEC® :
on the heart area
f
f
If you carry a pacemaker
f
f
If you are pregnant
f
f
during first 6 weeks after baby birth
f
f
In case of neuromuscular diseases
f
f
In case of hemorrhagic diseases
f
f
On the face and skull
f
f
cUSTOMER SUPPORT: +33 ( 0 ) 2 32 96 50 50
18
th
.
wARNINGS :
Keep away from children or people with mental disabilities.
f
f
The device contains small pieces that can be swallowed.
In order to prevent any risk of strangulation, please keep
f
f
the electrodes wires away from children.
If the device is not likely to be used for some time, the
f
f
batteries should be removed
Multisportpro is meant for personnal use. For reasons of
f
f
hygiene, the electrode should only be used on one person.
Do not use the device by driving or by working on a
f
f
machine.
Do not use this electro stimulation unit in a damp environ-
f
f
ment, a bathroom, a sauna, or close to electric machinery
connected to mains and earthed, or even piping earthed.
Use this device in a clean place (without dust nor dirt..).
f
f
The simultaneous connection of a patient to a high
f
f
frequency surgical unit may cause burns at the points of
contact of the stimulator's electrodes, and the stimulator
may be damaged.
Operation in close proximity of a short wave therapy unit
f
f
may cause instability in the stimulator's output power.
Do not, under any circumstances, attempt to open or
f
f
repair the Multisportpro yourself! Only have repairs carried
out by the customer service department.
RECOMMENDATIONS FOR MAINTAINING AND
CLEANING THE bELT :
IMPORTANT : NEVER wash in machine and under running
water. The dryer and ironing facilities are forbidden.
Do not dry-clean
f
f
Do not expose to bleach
f
f
Clean the belt using a cloth dampened with a detergent,
f
f
then wipe with a dry cloth.
Clean the unit with a tissue dampened with a disinfecting
f
f
solution (Mercryl type) diluted with waterat 1/10th.
V. TEchnIcaL SPEcIFIcaTIOnS
UNIT :
Class IIa medical device in accordance with Directive 93/42/CEE
Software version
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Type of current
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Nomber of programs
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Number of program variations (or phases)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Number of adjustable independent outputs
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maximum intensity on a 500 Ω load/channel
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Frequency range
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pulse width range
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power supply
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mains supply
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Unit dimensions and weight
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Automatic stop at end of each program
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Activation safety
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Temperature for stocking / Relative humidity
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Temperature for use / Relative humidity
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Caution !
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Guarantee of the unit / Lifecycle
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
bELT
References............................................................................................................
Size (unisex) ..........................................................................from 34 to 60
Length of the unfolded belt ..................................................... 126 cm
Extension lead ..................................................................................~35 cm
Fabric ................................................................................................. polyester
VI. GUaRanTEE
We provide a 24 month guarantee for products supplied by us, from the date of purchase.
f
f
The belt and adhesive electrodes are for personal use.
f
f
For hygiene reasons, they can therefore neither be returned nor exchanged.
f
f
The life expectancy of the adhesive electrodes is estimated at approximately 40 uses.
f
f
Never attempt to repair or modify the device yourself, or by a repairman not authorized by SPORT-ELEC®.
f
f
The guarantee becomes null and void if repairs have been carried out by non-authorized persons or if the replacement parts used
f
f
are not from the same manufacturer.
Similarly, only use compatible accessories manufactured by SPORT-ELEC®.
f
f
During the guarantee period we repair, free of charge, manufacturing defects or faulty parts, reserving the right to decide if
f
f
certain parts need repairing or replacing, or if the device itself must be replaced.
Any damage resulting from improper use (knocks, plugging into a direct current, wrong voltage, ...) does not fall under the
f
f
E
Dual phase
8 "SPORT" programs - 6 "HEALTH" programs
344
4
100 mA / 500 Ω / channel
0,25 to 160 Hz
70 to 360 μs
4,8 V - 1300 mAh - Ni-MH Rechargeable battery
THA239025 : 230VAC 50Hz 100mA - 9VAC 250mA
RSS1002-055110-W2B-B : 100-240VAC 50/60Hz 0.18A - 12VDC 250mA
The adapter is part of the equipment.
Only use mains power after reticulation by the power unit.
80 x 160 x 35 mm - 340 gr
Yes
100%
0°C to + 45°C / 10% to 90%
5-40°C / 20-65%
Read the manual thoroughly
2 years
Device in contact with the patient, excluding the heart
muscles.
Protected against the ingress of foreign object ≥12.5 mm &
against dripping water when tilted up to 15°.
These units are free of materials likely to have a negative
impact on the environment.
ADHESIVE ELECTRODES:
Class I medical device (Directive 93/42/CEE)
Adhesive, hydrophilic and hypoallergenic electrodes
References.................................................................................EASF110x71
Contact area ................................................................67.5 x 47 mm (x2)
Estimated life expectancy .......................................................... 40 uses
Please order new compatible adhesive electrodes on www.
sport-elec.com.
19