10.TROUBLESHOOTING
Problem
No weight displayed
with a load on the scale.
0.0 does not appear
before weighing.
One segment is illumi-
nated either continu-
ously or not at all.
The display
appears.
The display "bAtt"
appears.
The display "StOP"
appears.
The display "tEMP"
appears.
If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from
page 109.
CAUTION!
Bodily injury
If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 109, you may not use the device.
– Have servicing and repairs carried out
exclusively by seca service or by an
authorized service partner.
– Please see the section entitled
"Servicing/recalibration" on page 111.
Cause / Solution
The scale has no power supply.
- Check whether the scale is switched on.
A load was already on the scale before it was
switched on.
- Remove the load from the scale.
- Switch the scale off and back on again.
The corresponding point has a fault.
- Inform the service department.
Battery block power is running down.
- Recharge battery block as soon as possible.
Battery block discharged.
- Charge battery block.
Maximum load has been exceeded.
- Remove the load from the scale.
The ambient temperature of the scale is too high
or too low.
- Set up the scale in an ambient temperature
between +10 °C and +40 °C.
- Wait for around 15 minutes until the device
has adapted to the ambient temperature.
English
Troubleshooting • 109