OHAUS Guardian 7000 Serie Manual De Instrucciones página 15

Tabla de contenido

Publicidad

Guardian 7000 Series
4. To turn off the heater, press and hold the left knob
(C) until the unit beeps and the heater indicator (N)
disappears.
a)
The unit will beep once to confirm the heater
has been turned off.
The heater indicator (N) next to the "°C"
b)
symbol will disappear to indicate that the
heater is off.
CAUTION: This does not mean that the top
plate is safe to touch.
c)
The timer (T) will reset back to 00:00 or the
previous timer setting.
d)
Once the heater temperature (M) cools below
40°C, the hot top caution indicator (X) and
heater indicator bar (L) will disappear.
HEATING OPERATING TIPS
Overshoot:
The unit may overshoot the temperature up to 10°C before stabilizing at the setpoint. The three methods to
minimize overshoot are:
1.
Use the SmartRate™ Feature.
2.
Metal containers minimize overshoot. CAUTION! When heating metal containers on a ceramic top plate, it
is recommended to use the lowest temperature setting possible to limit thermal stress to the ceramic
top plate.
3.
If a glass vessel is used, anticipate overshoot. Start with a temperature setpoint 5 to 10°C below the desired
temperature. When the temperature stabilizes at this lower setting, increase the heater to the final
temperature. Overshoot is then reduced to about 1°C.
The temperature display on the units show the actual temperature of the heater, not the top plate or the sample.
The vessel's contents being heated may be at a lower temperature depending on the size and thermal conductivity
of the vessel. It may be beneficial to monitor the temperature of the vessel's contents and adjust the setpoint
temperature accordingly. If you need precise control, use the Ohaus External Temperature Probe.
Typical Time to Boil Water
The chart below is an example of an approximate time to boil for the specified amount of water in a specific vessel.
These values are only approximate and can vary from unit to unit. Values are based on 23°C water in an ambient
environment of 23°C.
Unit Size
Round Top
7×7
10×10
Heater Temp. Limit
380°C
500°C
500°C
Volume of Water
1L in 2L Beaker
1L in 2L Beaker
1L in 2L Beaker
EN-13
Typical Time to Boil
≈ 24 min
≈ 21 min
≈ 25 min

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido