Guardian 7000 Series
19. Repeat steps 8-15.
20. To clear a calibrated temperature point, return to
the Calibration settings menu. (Steps 1-3)
21. Rotate the left knob (C) to scroll to the 'CLR'
feature.
22. Briefly press the left knob (C) to enter the Clear
SPC menu.
If there are no stored SPC points, select 'BACK' to
return to previous screen.
23. Rotate the left knob (C) to scroll to the desired
SPC point.
SPC points are stored in ascending order by
temperature.
24. Briefly press the left knob (C) to select the point to
clear.
25. Rotate the left knob (C) to confirm selection:
'YES' – to clear the selected SPC point.
a)
'NO' – to return to Calibration settings menu.
b)
26. Briefly press the left knob (C) to confirm selection
and return to the Calibration settings menu.
27. To heat to a temperature setting with SPC, return
to the main operating screen.
28. Rotate the left knob (C) to scroll the heat setting
(O) to the desired temperature setting with SPC.
Notice that the 'SPC' icon (Y) appears.
EN-35