5. Operating parameter description

5.1 Input parameters

The addresses identified by * as well as "possible settings"
are only available dependent on the previously set up
parameter or the options released. The following listing
shows the maximum possibilities.
Once changes have been made to the setting up
parameters in the groups analogue input and
display/measurement range, please check the possible
effects these could have on other groups.
Parameter
Measurement range
Current input
Voltage input
Voltage input
(Only with "Universal
input" option)
Thermocouple
(Only with "Universal
input" option)
RTD
(Only with "Universal
input" option)
Units
Selection of engineering
units
(presettable)
Signal damping
Damping the input
signals
Decimal point
Selecting the decimal
point position
Display value 0%
Display value at 0%
sensor value
Setting up selections
RANG
4...20 mA, 0...20 mA
0...1 V, 0...10 V
±20 mV, ±50 mV, ±100 mV,
±200 mV, ±1 V, ±2 V, ±5 V,
±10 V
Typ T, Typ J, Typ K, Typ R,
Typ S, Typ B, Typ N, Typ U,
Typ L
Pt100, Pt500, Pt1000
Ni100
DIM1 DIM2 DIM3 DIM4
Value:
see table Chapter 6.2
There are four fields available
DIM1
= top left to
DIM4
= bottom right
DAMP
Value:
0 to 9999 s
DP
Selection:
0 to 3 decimal points
SCLO
Value:
-999 to 9999
The scale is preset on
temperature measurement!
Process display
LIME
Default
Actual
settings
settings
RANG
4-
20
%
DIM 1
DIM 2
DIM 3
DIM 4
0
DA
MP
000,0
DP
0,0
SC
LO
CH
AN
45

Publicidad