Date of next test:
Remarks: ......................................................
......................................................................
Operator: ......................................................
Signature: .....................................................
Step function measurement:
CHAUVIN ARNOUX C.A 6549
Instrument number: 700 016
Company: .....................................................
Address: .......................................................
......................................................................
Tel.: ...............................................................
Fax: ...............................................................
Email: ............................................................
Description: ..................................................
OBJECT: 01
TEST: 01
STEP FUNCTION TEST
Date:
Starting time:
Starting time:
Temperature:
Relative humidity:
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Step
Duration True def.
N°
h:m
voltage
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
1
00:10
1020 V
2
00:10
2043V
3
00:10
3060 V
4
00:10
3755 V
5
00:10
3237 V
R
V
R / (R* V) (ppm/v)
Capacitance
OBJECT: 01
TEST: 02
.... etc
Date of next test:
Remarks: ......................................................
......................................................................
Operator: ......................................................
Signature: .....................................................
../../.....
30.04.2009
14h55
00:15:30
23°C
. . . . %
Resistance
2.627 GOhm
2.411 GOhm
2.347 GOhm
2.182 GOhm
2.023 GOhm
604 GOhm
4000 V
-57 ppm
100 nF
../../.....
102