After the measurement
DAR
234.5 MΩ
5007 V
224.6 pA
Elapsed Time 00:01:00
DAR (30s/60s)
PI (1.0m/10m)
Capacitance
220 nF
Available information:
First screen
Insulation resistance
Measured voltage
Measured current
Elapsed test duration
DAR, PI, capacitance
 PI mode
Same as DAR mode except:
PI instead of DAR at the top left of the display unit
Remaining Time = 10 min
After the measurement: display of DAR and PI.
4.2.2. GRAPH SECONDARY FUNCTION
This function is used to display the insulation resistance versus measurement time curve after a time-
limited measurement (Timed Run or Timed Run + DD).
This curve is plotted from the samples recorded during the measurement. The
can be used to move along the curve to display the exact values of each sample.
GRAPH
5078 V
2428.5 MΩ
00:02:30
3000
2500
2000
1500
1000
500
0
0:30 1:00 1:30 2:30 3:00
DAR
FIXED VOLTAGE
5000 V
Test Run Time 00:01:00
2.24
Input voltage
---
Frequency
Input current
Date: 24.06.2008
Press on DISPLAY
Selected test voltage
Programmed test duration
Input voltage
Frequency
Input current (DC only)
Date, time
83
DAR
0.1 V
0.0 Hz
0.1 V AC
0.0 Hz
0
250
100
24.6 pA
Time: 15:32
2
nd
press on DISPLAY
Input voltage
Frequency
Input current (DC only)
Voltage bargraph
,
24.6 pA
AC
V
500
750
1000
,
and
keys