Advanced functions
Pressing
swaps between the lower and upper limits.
Pressing
exits limit mode and saves any changes.
Message <LIMIT
(low sound)
lower
limit value
- full scale
value
Message ><LIMIT (sound)
upper
limit value
- full scale
value
Fuse protection in Ω-function:
When in continuity mode fuse damage will occur if the test leads are applied to a
live circuit. The
symbol appears to indicate the fuse should be replaced.
Continuity testing can be carried out without any risk of fuse damage in the normal
resistance mode by setting the Lo LIMIT to the appropriate value and switching the
Hi LIMIT off. In this mode the continuity test is slower than the normal
function. Test time can be by switching off the auto range function during power on
with the LIMIT key.
Message >LIMIT
(high sound)
upper
limit value
+ full scale
lower
limit value
+ full scale
continuity
- 21 -
value
value