Operation
4.9
Common user tasks
4.9.1
Replacing the batteries / rechargeable batteries
WARNING
!
Explosion hazard!
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in a hazardous area.
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
Alkaline batteries – T3 – (non rechargeable!)
Panasonic LR6 Powerline
1
Varta Type 4106
1
Varta Type 4006
Alkaline batteries – T4 – (non rechargeable!)
Duracell Procell MN1500
NiMHy rechargeable batteries – T3 – (rechargeable)
1
GP 180AAHC
temperature.
Charge NiMH power packs (type HBT 0000) or T4 HC (type
HBT 0100) with the appropriate Dräger charger. Charge
NiMH rechargeable batteries for battery holder ABT 0100 in
accordance with the manufacturer's specifications. Ambient
temperature during the charging process: 0 to +40 °C.
1)
Not subject to BVS10 ATEX E 080X and PFG 10 G 001X performance approval.
1. Switch off instrument: Press and hold OK-key and [+]-key
simultaneously.
2. Loosen the screw on the power pack and remove the power pack.
For battery holder (order no. 83 22 237): replace alkaline batteries or
respectively NiMH rechargeable batteries. Ensure correct polarity.
22
(power one) or
(industrial)
1
(1800 mAh) max. 40 °C ambient
For NiMH power pack T4 (type HBT 0000) / T4 HC (type HBT
0100): Replace entire power pack.
3. Insert the power pack into the instrument and tighten the screw,
the instrument switches on automatically.
4.9.2
Charging the instrument with NiMH power pack T4 (type
HBT 0000)/ T4 HC (type HBT 0100)
WARNING
!
Explosion hazard!
Do not charge underground or in explosion hazard areas!
The chargers are not designed in accordance with the
regulations for fire damp and explosion protection.
Charge NiMH power pack T4 (type HBT 0000) or T4 HC (type
HBT 0100) with the appropriate Dräger charger. Ambient
temperature during the charging process: 0 to +40 °C.
Even if the instrument is not used, Dräger recommends that you store
the instrument in the charger cradle!
Insert the switched off instrument into the charger.
Display LED on the charger:
Charge
Malfunction
Full
To conserve the lifetime of the batteries, a charge is only carried out in a
temperature range of 5 to 35 °C. When this temperature range is left, the
charging process is automatically interrupted and automatically continued
after the temperature range has been reached again. The charging time
is typically 4 hours. A new NiMH power pack reaches its full capacity after
three complete charge/discharge cycles. Never store the instrument for
extended periods without being connected to a power source (maximum
of 2 months) because the internal buffer battery will drain.