Calling the Quick Menu
In measuring mode, press the [+] key three times.
If functions in the quick menu are activated using the PC software
"Dräger CC-Vision", you can select these functions using the
[+]-key. If no functions have been activated in the quick menu,
the instrument remains in measuring mode.
Possible functions:
1. Bump test
2. Fresh air calibration
3. Delete peak values
Press the OK key to activate the selected function.
Press the [+] key to cancel the active function and to switch to
measuring mode.
If no key is pressed for 60 seconds, the instrument returns
automatically to measuring mode.
Dräger X-am 2500
4.9

Common user tasks

4.9.1
Replacing the batteries / rechargeable batteries
WARNING
!
Explosion hazard!
Do not throw used batteries into fire or try to open them by
force.
Do not replace or charge batteries in a hazardous area.
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
Alkaline batteries – T3 – (non rechargeable!)
Panasonic LR6 Powerline
1
Varta Type 4106
1
Varta Type 4006
Alkaline batteries – T4 – (non rechargeable!)
Duracell Procell MN1500
NiMH rechargeable batteries – T3 – (rechargeable)
1
GP 180AAHC
(1800 mAh) max. 40 °C ambient
temperature.
Charge NiMH power packs T4 (type HBT 0000) or T4 HC
(type HBT 0100) with the appropriate Dräger charger.
Charge NiMH single cells for ABT 0100 battery holder as
directed by the manufacturer. Ambient temperature during
charging: 0 to +40 °C.
1)
Not part of the BVS10 ATEX E 080X and PFG 10 G 001X technical suitability test.
Operation