Operation
4.9

Common user tasks

4.9.1
Replacing the batteries / rechargeable batteries
WARNING
!
Explosion hazard! To reduce the risk of ignition of a flammable
or explosive atmosphere, strictly adhere to the following
warning statements:
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in potentially explosive
areas.
Do not use new batteries with used batteries, and do not mix
batteries from different manufacturers or of different types.
Remove batteries before maintenance work.
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
Alkaline batteries – T3 – (non rechargeable!)
Panasonic LR6 Powerline
Varta Type 4106 (
1
Varta Type 4006
Alkaline batteries – T4 – (non rechargeable!)
Duracell Procell MN1500
NiMHy rechargeable batteries – T3 – (rechargeable)
1
GP 180AAHC
temperature.
Only charge NiMH power packs T4 (type HBT 0000) or T4 HC
(type HBT 0100) with the appropriate Dräger charger. Charge
NiMH single cells for ABT 0100 battery holder as directed by
the manufacturer. Ambient temperature during charging:
0 to +40 °C.
1)
Not part of the BVS10 ATEX E 080X and PFG 10 G 001X technical suitability test.
30
1
power one) or
(industrial)
1
(1800 mAh) max. 40 °C ambient
1. Switching off the instrument: Press and hold OK key and [+] key
simultaneously.
2. Loosen the screw on the power pack and remove the power pack.
Battery holder (order no. 83 22 237): Replace alkaline batteries or
NiMHy rechargeable batteries. Ensure correct polarity.
NiMH power pack T4 (Type HBT 0000) / T4 HC (Type HBT 0100):
replace complete power pack.
3. Insert the power pack into the instrument and tighten the screw,
the instrument switches on automatically.
4.9.2
Charge instrument with NiMH power pack T4
(Type HBT 0000) / T4 HC (Type HBT 0100)
WARNING
!
Explosion hazard! To reduce the risk of ignition of a flammable
or explosive atmosphere, strictly adhere to the following
warning statements:
Do not charge underground or in explosion hazard areas!
The chargers are not designed in accordance with the
regulations for fire damp and explosion protection.
Charge NiMH power packs T4 (type HBT 0000) or T4 HC
(type HBT 0100) with the appropriate Dräger charger.
Ambient temperature during charging: 0 to +40 °C.
Insert the switched off instrument into the charger module.
Display LED on the charger module:
Charge
Malfunction
Full
Dräger X-am 2500