NOTE– The EMC tables and other guidelines provide information to the customer or user that is essential in
determining the suitability of the Equipment or System for the Electromagnetic Environment of use, and in
managing the Electromagnetic Environment of use to permit the Equipment or System to perform its intended
use without disturbing other Equipment and Systems or non-medical electrical equipment.
guidance and Manufacturer's declaration – Emissions all Equipment and
Systems
This device is intended for use in the electromagnetic environment specified below. The customer or user of
this device should assure that it is used in such an environment.
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonics
IEC 61000-3-2
Flicker
IEC 61000-3-3
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
Transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
A-DV5M-1
Compliance
Group 1
Class B
Radiated and
Conducted Emissions
Class A
Complies
IEC 60601 Test
Level
±6kV contact
±8kV air
±2kV on AC Mains
±1kV Differential
±2kV Common
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
Electromagnetic Enforcement – Guidance
This device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
This device is suitable for use in all establishments including
domestic, and those directly connected to the public low-
voltage power supply network that supplies buildings used
for domestic purposes.
Compliance Level
±6kV contact
±8kV air
±2kV on AC Mains
±1kV Differential
±2kV Common
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
Electromagnetic Environment -
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are synthetic,
the relative humidity should be at
least 30%.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of this device
requires continued operation during
power mains interruptions, it is
recommended that the device be
powered from an uninterruptible
power supply or battery.
EN - 11